Máy sinh hóa,tự động,AT240,AJR Medical Inc,Mỹ

AUTOMATIC CHEMISTRY ANALYZER: MODEL- AT240

TECHNICAL SPECIFICATIONS
Overall  function
 - Equipment: Fully automated, discrete, random-access, STAT priority
 - Through-put: 240  tests / hour
 - Test principle: colorimetry, turbidimetry
 - Analysis principle: 
   + End-point, kinetics,  fixed  time, …etc
   + Support single/ double wavelength and 1-2 multiple reagent item
   + Linear and non-linear calibration

Sample  reagent unit
- Sample  reagent position: 67 sample &  reagent positions shared  in a disk. Continuously cooling  to keep 5~15ºC within 24 hours
- Sample volume: 3~35µL, 0.1µL stepping
- Sampling  technology: liquid  level detection, clot detection and collision detection
- Sampling probe: 1, with the function of liquid level detection
- Crossing  rate: automatic warm water cleaning. Carrying  rate  0.1%
- Automatic sample dilution: 2~115  times

Reaction Handling
- Reaction cuvette:120, optical plastic cup, optical diameter  is 6mm
- Total volume of  reaction  liquid: 150 µl~550µl
- Reaction  temperature: 37ºC ± 0.1ºC 
- Reaction disk constant temperature: Circulating water
- Mixed needle: 1, after joining reagent, blending immediately
- Reaction cuvette cleaning: 8 stops 12 steps by warm water  rinsing
- Wastewater treatment: with the function of concentrated waste liquid level alarming

Optical system

- Light source: 20W/12V halogen  lamps. Life is 2000 hours
- Monochromator: Grating photometry
- Photoelectron  road: after spectrophotometry
- Wave  length: 340, 380, 405, 450, 480, 505, 546, 570, 600, 660, 700, 750  (or 800nm)
- Detector: photodiode LED array
- OD  linear  range: 0~3.3Abs

Calibration & QC
- Calibration method: 1 point  linear, 2 point  linear, multi point  linear, non-linear method
- Calibration  tracking: Automatic description calibration K-value  trends
- QC method: Real-time QC, days QC and day QC
- Out of control processing: Alarming  for out of control sample,  record  lost control  reason

Others:
- Volume: 998 x 752 x 517 cm
- Weight: 120 kg
- Power supply: voltage AC 220V ± 22V, 50/ 60 Hz,
- Power: 650VA