Monitor

Monitor,Deluxe-140 multi-parameter monito,Heal Force,TQ

Monitor,Deluxe-140 multi-parameter monito,Heal Force,TQ

 Applications

Our Deluxe-140 multi-parameter monitor is suitable for monitoring parameters of ECG, RESP, NIBP, SpO2 and TEMP. Other optional monitoring applications are also available, such as mainstream, sidestream EtCO2 and 1 or 2-Channel IBP.

Features
1. Deluxe-140 multi-parameter monitor comes with a 14.1"color TFT display. Up to 9 curves can be simultaneously displayed.
2. It has accessories for adult, neonatal and pediatric patients. 
3. The multi-parameter monitor provides visual and audible alarms with individual alarm setting.
4. Maximum 480 hours trend graphs enable nearly all parameters and trends to be listed for further analysis. 
5. The patient monitor allows storage of up to 1000 groups of NIBP data.
6. It can be wall mounted or mounted with rolling stand. 
7. Network capability with PC-1000 PRO central monitoring system is available.



Specifications of Deluxe-140 Multi-Parameter Monitor 
ECGHeart rate display range15bpm ~ 350bpm
Heart rate display accuracy± 1% or ± 2bpm, whichever is greater
Sweeping speed12.5mm/s, 25mm/s, 50mm/s
Sensitivity selection×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV
CMRR≥ 89dB
RESPMeasuring range0rpm~120rpm
Accuracy±5% or ±2 rpm
Alarm limits setting range0rpm~120rpm
TEMPMeasuring range25.0℃ ~45.0℃
Measuring accuracy±0.2℃
Responding time≤150s
NIBPMeasuring methodOscillometric Technique
Measuring range0mmHg~300mmHg
AccuracyMaximal mean difference±5 mmHg
Maximal standard deviation8 mmHg
Measurement rangeAdultSYS 40 ~ 255 mmHg
MAP 20 ~ 215 mmHg
DIA 10 ~ 195 mmHg
Pediatric patientSYS 40 ~ 200 mmHg
MAP 20 ~ 165 mmHg
DIA 10 ~ 150 mmHg
NeonateSYS 40 ~ 135 mmHg
MAP 20 ~ 110 mmHg
DIA 10 ~ 95 mmHg
SpO2Measuring range35%~100%
Low perfusion capability0.4%~5%
AccuracyNot greater than 3% for SpOrange from 70% to 100%
PRMeasuring range30bpm~240bpm
Measuring accuracy±2bpm or ±2%, whichever is greater