Phụ kiện máy chạy thận nhân tạo

Máy rửa, quả lọc thận,JH-FYAD,Jihua,TQ

Máy rửa, quả lọc thận,JH-FYAD,Jihua,TQ

 

JH-FYAD Dialyzer reprocessing machine

 

 Main technical parameter:

 Exterior size: 480mm×430mm×560mm
    ·Weight:About 35kg
    ·circumstance temperature: 5 ~ 40 
    ·Relative humidity: 10%--80%
 Dialyzer test:
    ·Pressure test:put 290mmHg± 

10mmHg pressure to one side of blood chamber .
pressurerelease of dialyzer should be less than:
       Low flux dialyzer :0.83mmhg/sec
       High flux dialyzer :25mmhg/sec
  ·Capacity test: Scope:0170ml, error: ±3%
 Water required: Water temperature:10~30
    ·Water quality required: RO water reach the
       standard 5.1.2 of<<operation regular of dialyzer
       reprocessing equipment>>of Sanitary Ministry
    ·Water input pressure:0.5MPa~0.35MPa
    ·Water input flux:>1600ml/min

Rinse time:
   ·Rinsing time:2min/time
   ·Disinfection time: 7min/time
  Reprocessing dialyzer time: 11min/time
    ·Water amount:about 8L/dialyzer
    ·Power supply: AC220V±10%
    Disinfection solution:common disinfectant,ie,
    peracetic acid renalin and other special
    disinfectant.
    Proposed use: 12% peracetic acid,100% renalin
                          (alternative)
Print parameter:
     

·English: 6*8 lattice print rate:1sec./row.
·Chinese:12*12 lattice.

Function

 It could clean and sterilize different high and low flux dialyzer,and inspect blood capacity automatically as well as membranebroken experiment and pressure inspection,it can clean,sterilize and inspect two dialyzers at the same time.With the printer 

function,it can print the inspected data out and stick on cleaned dialyzer to judge if the dialyzer can continue to use or not.

Function setting

 Switch on self-test function.
 Information memory of all aspect of dialyzer which facilitate to operate.
 Entirely new touch screen.
 Spare pump counter-rinse,to reach the best effect.
 Automatically inspect the precision of blood capacity sensor.
 Automatically inspect the precision of pressure sensor.
 Automatically inspect if exterior connector link tightly or not.
 Be capable to handle any kinds of hollow fiber dialyzer.
    ·(HD)Common dialyzer
    ·(HDF)High efficiency and high-flux of hemodiafiltration dialyzer
    ·(HF)hemofiltration dialyzer
 Humanity operation interface
 Design for saving water and low inflow pressure to unload center water supply.
 Be provided with data saving when power cut.
 Professional maintenance interface which can inspect if all valves and pumps work normally or not .
 Be provided with original creation function.
 With a special detector to inspect disinfection solution concentration.which assure the required disinfection concentration 

filled with dialyzer to aviod the disinfection failure caused by manmade which occur medical accident.
 Be provided with alarm for block or fold of pipe which assure liquid filling with dialyzer during
disinfecting.With a highly security and complete clean.
 Confecting disinfection liquid automatically by system as well as rinsing and filling up dialyzer.