DANH MỤC SẢN PHẨM

Tìm kiếm nâng cao

Hỗ trợ trực tuyến

Bình chọn

  • Bạn biết về trang web qua nguồn nào ?
  • Internet
  • Hội chợ,Triển lãm
  • Báo chí
  • Qua tổng đài 1080
  • Người thân,Bạn bè
  • Tờ rơi,Poster
  •  

Tỉ giá ngoại tệ

Đối tác

Thống kê

Số lượt truy cập 1541260   lần

54   khách hàng đang online

Máy tán sỏi ngoài cơ thể

Máy tán sỏi ngoài cơ thể, ESWL-108,Hyde,TQ

Máy tán sỏi ngoài cơ thể, ESWL-108,Hyde,TQ

Product Description:

1. Application                                                                                          

1.1. Urological calculi

1.2. Urological operation table

 

2. Working condition

2.1. Environmental temperature: 1535℃

2.2. Relative humid: 4585%

2.3. Atmospheric pressure: 86106kPa

2.4. Power requirements: AC 220V  50/60±1Hz,  PW≤3kW

2.5. Water requirements: Dematerialized water

2.6. Room requirements:

2.6.1. Main room 3m×3m×2.5m

 

3. Shock wave generator: Electromagnetic & Electrohydraulic

3.1. Discharging range: EM: 10~16kV  EH: 5~11kV

3.2. Capacity storage: EM: 46~104J  

3.3. Parameter of second focus of shock wave

3.3.1. Shock wave front 0.5mS

3.3.2. Shock wave width 1.0mS

3.3.3. Focal spot area: 7×7×25 mm

3.3.4. Peak value of second focus of shock wave: 20~50mPa

3.3.5. Length of second focus: 140mm

4. Localization system

4.1. B-ultrasound localization

4.1.1. Probe extension

4.1.1.1. The localization error may be caused by the probe makes linear movement relative to the focus is within ±2mm

4.1.1.2. The localization error may be caused by the probe makes annular movement relative to the focus is within ±2mm

4.1.2. The measurement error between the probe surface and the second focus is within ±2mm

 

5. Therapy table

5.1. 3-D mechanical movement: X100mm, Y100mm, Z100mm

5.2. Minimum adjustment of 3-D movement of the table 1mm

5.3. Maximum capacity of the table 135kG

 

6. Operating system

6.1. PLC and microcontroller control module

6.2. Energy level adjustment with level

6.3. Free adjustment of discharging frequency within 0.32 times per second

6.4. Automatically search and indicate malfunction

 

7. Water dealing system

7.1. Automatic degassing and recycling system

7.2. Automatic constant temperature device

7.3. Overheat alarm and protection device

 

8. Main configuration of the lithotripter (reference)

8.1. Main unit

8.2. Therapy table

8.3 Console

8.4 Electromagnetic shock wave source (EM pan, lens, HV unit and cushion)

8.5. B-ultrasound localization device (without B-ultrasound)

8.6. Auxiliaries