Monitor

Monitor DRE Waveline Pro Multi-Function

Monitor  DRE  Waveline Pro Multi-Function

Dre-Mỹ

DRE Waveline Pro Multi-Function Patient Monitor


Product Description

 

Superior flexibility and performance with a premium flat panel display

The DRE Waveline Pro displays as many as eight colorful waveforms on a crisp 15” high resolution display. It is lightweight, portable and measures multiple parameters, including ECG (3 or 5 lead) with ST segment monitoring, resp., dual temp., SpO2 and NIBP. Configurations are available that measure capnography, dual channel IBP, and anesthetic agents (five agents, N2O, FiO2, EtCO2), and that have a built-in thermal printer.

FEATURES

  • Displays as many as eight waveforms
  • Available with automatic agent ID
  • Multi-lead simultaneous ECG monitoring
  • Automatically set alarm limits
  • Graphical and tabular trending
  • Color-coded alarms
  • Patient data entry
  • Battery backup (two removable batteries)
  • Networking capability
  • Large (15”), high resolution flat panel display
  • Available with a built-in thermal printer

  

SPECIFICATIONS

Physical Dimensions & Weight

  • Base Unit: 12.1”/309 mm (H) x 14.4”/365 mm (W) x 6.3”/159 mm (D)
  • Weight: 17.6 lbs/8.0 kgs (including anesthetic agents module and batteries)

Application

  • Neonatal, pediatric and adult patients

Performance Specifications

  • Display: 15” color TFT, Resolution 1024 x 768
  • Trace: 8 waveforms
  • Indicators:
    Alarm indicator
    Power indicator
    QRS beep and alarm sound
  • Trend time: 1 – 48 hours
  • Recorder:
    Built-in, thermal array, 2 channels
    Record width: 1.9”/48 mm
    Recorder paper: 2”/50 mm
    Record speed: 25 mm/s, 50 mm/s

ECG

  • Input: 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice: I, II, III, aVR, aVF, aVL, V, TEST
  • Gain Choice : x0.5, x1, x2, x4
  • Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
  • ECG Waveforms: 7 channels
  • Penetration Voltage: 4000VAC 50/60Hz
  • Sweep Speed: 12.5, 25 and 50 mm/s
  • HR Display Range: 30 – 300bpm
  • Accuracy: ±1 bpm or ±1%, whichever is greater
  • Alarm Limit Range Setting: Upper limit 100 – 200 bpm; Lower limit 30 – 100 bpm

RESP

  • Measure Method: RA-LL impedance
  • Range: 0 – 120 rpm
  • Accuracy: ±3 rpm
  • Alarm Limit Setting: Upper limit 6 – 120 rpm: Lower limit 3 – 120 rpm
  • Sweep Speed: 12.5 and 25 mm/s

TEMP

  • Range: 25° – 50° (C)
  • Accuracy:
    ± 0.2° C (25.0° – 34.9° C)
    ± 0.1° C (35.0° – 39.9° C)
    ± 0.2° C (40.0° – 44.9° C)
    ± 0.3° C (45.0° – 50.0° C)
  • Display Resolution: 0.1° C
  • Alarm Limit Setting: Upper limit 0° – 50° C, lower limit 0° – 50° C
  • Channel: 2 channels

SpO2

  • ASpO2: Anti-motion SpO2
  • SpO2% Range: 0 – 100%
  • SpO2 Accuracy: ±2% (70 – 100%, non-motion) ±3% (70 – 100%, motion)
  • Pulse Rate Range: 30 – 250 bpm
  • Pulse Rate Accuracy: ±2 bpm (non-motion) ±3 bpm (motion)
  • Alarm Limit Setting: Upper limit 70 – 100%, lower limit 70 – 100%
  • SpO2 Probe:
    Red light LED wavelength – 660nm±5nm
    Infrared light LED wavelength – 940nm±10nm

NIBP

  • Measuring Technology: Automatic oscillating measurement
  • Cuff Inflating: <30s (0 – 300 mmHg, standard adult cuff)
  • Measuring Period: AVE<40s
  • Mode: Manual, Auto
  • Measuring Interval in AUTO Mode: 2 min – 4 hrs
  • Pulse Rate Range: 30 – 250 (bpm)
  • Measuring Range:
    Adult/Pediatric Mode
    - SYS: 40 – 250 (mmHg) DIA :15 – 200 (mmHg)
    Neonatal Mode
    - SYS: 40 – 135 (mmHg) DIA : 15 – 100 (mmHg)
  • Resolution: 1mmHg
  • Accuracy:
    Maximum Mean error: ±5mmHg
    Maximum Standard deviation: 8mmHg
  • Overpressure Protection:
    Adult Mode: 300(mmHg)
    Neonatal Mode: 160 (mmHg)
  • Alarm Limit Setting:
    - SYS: 50 – 240 mmHg
    - DIA: 15 – 180 mmHg

IBP

  • Measurement Range: -50 – 300mmHg
  • Channel: 2 channels
  • Pressure Transducer: Sensitivity, 5μV/V/mmHg
  • Impedance Range: 300 – 3000Ω
  • Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
  • Unit: mmHg/kPa selectable
  • Resolution: 1 mmHg
  • Accuracy: ±1 mmHg or ±2%, whichever is greater
  • Alarm Range: -10 – 300 mmHg

EtCO2 (Oridion)

  • CO2 Measurement Range: 0 – 99 mmHg
  • Accuracy:
    ±2 mmHg (0 – 38 mmHg)
    39-99 mmHg ±5% of reading + 0.08% for every 1 mmHg (above 38 mmHg)
  • Sampling Rate: 50 ml/min
  • Initialization Time: 30 seconds (typical), reaches ±5% steady-state accuracy within 3 minutes
  • Respiration Rate: 0 – 150 breaths/min
  • Mode: Adult, neonate

Anesthetic Agents

  • Method: Infrared absorption
  • Gas Sorts: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, O2 (optional Automatic Agent ID)
  • Measurement Range:
    Halothane, Isoflurane: 0 – 8.5%
    Enflurane, Sevoflurane: 0 – 10%
    Desflurane: 0 – 20%
    CO2: 0 – 10%
    N2O: 0 – 100%
    O2: 0 – 100%
  • Bias:
    Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane: ±(0.15 Vol% + 15% rel.)
    CO2: ±(0.5 Vol% + 12% rel.)
    N2O: ± (2 Vol% + 8% rel.)
  • O2: ±3 Vol%

Networking

  • Industry standard 802.11b/g wireless network

Power

  • Source: External AC power or internal battery
  • AC Power: 100 – 240VAC, 50/60Hz, 150VA
  • Battery: Built-in and rechargeable
  • Charge Time: 4 hours

Environmental Specifications

  • Temperature:
    Operating: 5° – 40° C
    Storage: -20° – 65° C
  • Humidity range:
    Operating: ≤80 %
    Storage: ≤80 %