Nồi hấp

Máy hấp tiệt khuẩn 80L Model: ASVE/A Hãng sản xuất: SMS Xuất xứ: Ba Lan

STEAM STERILIZERS

ASVE/A vertical sterilizer capacity 80 litres

The ASVE/A sterilizer is intended for steam sterilization of glass, infusion liquids, microbiological cultures and nutrients, surgical instruments, operation linen, dressing materials, rubber products, metal and plastic dishes. The sterilizer is used in laboratories, hospitals, clinics, outpatient clinics and pharmacies in a great way.

Product description:
cylindrical, vertical positioned sterilizing chamber
7 sterilizing programs which allow modification of temperature and duration of the exposure, drying and cool-down phases, and 2 service programs (Bowie-Dick and air-tightness test); also user-definable program is available, in which any configuration of the individual phases can be obtained
built-in steam generator
microprocessor control; control panel with a keypad, LED indicators and LCD screen for displaying messages while setting up or operating the sterilizer
built-in sterilization parameter printer (AD model)
stainless steel chamber, steam generator and housing, rust-proof fittings

Optional solutions:
sterilizer to be connected to an external steam source, which ensures pressure 0.25 MPa and
12 kg/h flow (average consumption 3.5 kg/cycle)
HEPA additional exhaust air filter
software for monitoring, recording and archiving parameters of individual sterilizing cycles
closed-circuit water tank
additional temperature sensor in chamber for sterilizing liquids
sterilizer mounted on wheels
stainless steel fittings

Technical specifications:

Type
Unit
ASVE/A
Total sterilizing chamber capacity
litre
80
Chamber dimensions
mm
Ø400×600
Overall dimensions (width, height, depth)
mm
872×1110×740
Weight
kg
240
Sterilization temperature
°C
110÷138
Vacuum drying

ejector-vacuum and spinning water pumps
Water for vacuum pump supply – max consumption (supply pressure 0,1÷0,6 MPa)
l/min
25
Water for steam generator supply (recommended water hardness max 7°n, supply pressure 0,1÷0,6 MPa)
max consumption
avg. consumption


l/min
l/cycle


0,21
3
Electrical supply 3×230/400 V, 50 Hz, installed power
kW
9,5

Additional accessories:
wire basket dimensions Ø360×275 mm (KSV 1/2)
wire basket dimensions Ø360×140 mm (KSV 1/4)
solid basket dimensions Ø360×275 mm (PSV 1/2)