Máy đo Oxy trong máu và theo dõi nhịp tim

Máy đo oxy trong máu,Jerry-II(AA),Shanxi Jerry,TQ

Máy đo oxy trong máu,Jerry-II(AA),Shanxi Jerry,TQ

 JERRY-II(AA) Pulse Oximeter

 

Product name:

JERRY-II(AA) Pulse Oximeter

  
Characteristic:

 

The monitor is a non-invasive, handheld patient monitor. It operates only on Common 1.5V AA alkaline or rechargeable battery power. It is compact, small, light, and easy for learning and handling. It is suitable for monitoring adult, pediatric, infant and neonate patients. It is widely used in the hospital’s operation room, ICU, clinic section office, out-patient department, sickroom, emergency treatment, and the recovery and health care organizations, or in the family nursing and in the process of transporting patients. It can also apply to animals selectively.
Parameters measured by the monitor include: arterial oxygen saturation (SpO2), pulse rate (PR), pulse strength. The monitor measures these parameters through a SpO2 sensor and displays them on the LED screen after certain further processing.
The monitor is operated and controlled by the buttons on the front panel. It adopts LED in displaying measurements and in supplementary status indication.
The monitor has the following functions:
■ Measuring: Display of SpO2, PR and pulse strength.
■ Prompting: Battery capacity.
■ Alarm: Judgment of exceeding the limits, Audio and visual alarm.
■ Power saving: Automatic shutdown when no monitoring.
 
H.S.Code: 90189090
Net Weight: 0.13kg 
Gross Weight: 0.8kg
Package Size: 210x160x70mm
Color: Yellow(with JERRY logo), Blue(without JERRY logo) 
 
Standard accessories: one SpO2 adult finger probe
 Technical specification
General
Monitoring Parameters:SpO2、PR、Pulse strength
Signal sockets:SpO2 socket
Display screen            
Type:Red LED
Display area:56mm×36mm
Size:120mm×63mm×32mm
Weight:125g(not include sensor and battery) 
Features
Audio and visual alarm:  The speaker sounds the alarm and the parameter’s measurement flashes on the screen.
Prompting: Indicator of battery capacity.
Power saving: The monitor will automatic shut down when no monitoring for 3 minutes.
Electrical specifications
Working Voltage:D.C. 2.8V~D.C. 5.4V.
Battery 
Type:Three common 1.5V AA alkaline or rechargeable batteries
Run Time:7-hour continuous operation with a new battery (environment
temperature is 25℃)
Environment
Temperature
Operation:0 ~ 45 (°C)   
Transportation and storage:-20 ~ 60 (°C)
Humidity
Operation:15% - 95 %( noncondensing)  
Transportation and storage:10% - 95 % (noncondensing)  
Altitude
Operation:86 KPa ~106 KPa
Transportation and storage:50KPa~106 KPa
Parameter Specifications
SpO2
Patient:Adult, pediatric, infant, neonate, animal
Range:35 - 99%
Resolution:1 %
Accuracy:±2 % (70% - 99%), unspecified (0% - 69%)
PR
Range:30~250bpm
Resolution:1bpm
Accuracy:±2bpm