Máy truyền dịch

Máy truyền dịch Top-3300- Nhật

点击放大

In the busy world of nursing,
information has to be prioritized.

First and foremost, one needs to know whether the pump is working properly.
The TOP-3300 continuously conveys this crucial bit of information with the lighted status indicators provided at the top of the unit. Also, the panel face is inclined upward, making it possible to check the display and work the keys on the panel without having to bend over.
The TOP-3300 - a new infusion pump which fully meets the conditions of use even on general hospital wards, where the level of care cannot be as intensive.
Lighted Status Indicators
LED status indicators at the top of the unit allow the pump operation status to be checked even from the back of the unit or at a distance.
 
Inclined Panel for Easy Visual Confirmation
Control panel is inclined upward, making it easy to check the display and allowing for smooth key operation from a standing position.
 
Broad Range of Delivery Rate Settings
The delivery rate can be set up to a maximum of 999 ml/h (for infusion in the H mode using a TOP IV set, H series), and can be adjusted in 0.1 ml/h increments up to 99.9ml/h.
 
Convenient Battery Indicator
A 3-level indicator displays the charging status and remaining capacity of the internal battery, so that the operator knows when to get spare batteries of switch over to AC power.
 
Battery Replacement is Easy
The batteries can be easily replaced simply by opening the battery cover at the bottom of the unit.
 
Improved Drop Sensor
The drop sensor which monitors drip infusion has been upgraded to a thinner design than in earlier TOP infusion pumps, enabling drops to be counted correctly even when the sensor is tilted at up 20 degrees.
 
Occlusion Alarm-Triggering Pressure is Adjustable
The load pressure detection level for alerting the user to occlusion of the infusion line can be switched as required.
 
Memory Function
When the power is turned off, the unit stores and retains the most recent delivery rate, volume limit, and total volume infused.
 
Clears EMC Standards
The pump conforms to EU standards for electromagnetic compatibility (EMC), which require that devices not be affected by electromagnetic fields form external sources and not emit electromagnetic waves sufficient to affect the operation of other electronic devices.
 
Specifications
Pump type
Peristaltic finger
 
Applicable IV sets
TOP IV sets H-Series (TOP-H) for infusion pumps, and other common IV sets (15, 19, 20, and 60 drops/ml)
 
Delivery rate range
1.0-999ml/h (TOP_H series dedicated IV set)
1.0-400ml/h (General-purpose 15 drops/ml IV set)
1.0-380ml/h (General-purpose 19 drops/ml IV set)
1.0-360ml/h (General-purpose 20 drops/ml IV set)
1.0-150ml/h (General-purpose 60 drops/ml IV set)
1.0-99.9ml/h, in 0.1 ml/h increments; above 100ml/h, in 1 ml/h increments
 
Volume limit range
1 to 9,999ml, in 1 ml increments
 
Infusion accuracy
±5% when TOP-H sets are used
±10% when other common IV sets are used
 
Alarms

Empty container, Occlusion, Door open, Air-in-line, Low battery, infusion completed, Power source switched, Operator error, and other device malfunctions

 
Air bubble detection
Detects air bubbles 50 L or larger in size
 
Occlusion detection pressure

Can be set to any of 4 levels when TOP-H sets are used:
     P1-4: occlusion not detected
     P1-3: 160.0 kPa (1,200 mmHg/1.63 kgf/cm2)
     P1-2: 120.0 kPa (900 mmHg/1.22 kgf/cm2)
     P1-1: 80.0 kPa (600 mmHg/0.82 gf/cm2)
     P1-0: 40.0 kPa (300 mmHg/0.41 kgf/cm2)

Can be set to either of 2 levels when other common IV sets are used:
     P2-1: 160.0 kPa (1,200 mmHg/1.63 kgf/cm2)
     P2-0: 80.0 kPa (600 mmHg/ 0.82 kgf/cm2)
 
Air bubble detection
Detects air bubbles 50 L or larger in size
 
Special functions
Automatic shutoff: When the infusion pump is operating on battery power and has been left for 3 minutes with the infusion stopped or in an alarm state, the buzzer sounds. After the buzzer has been left sounding for 3 minutes, the power turns off automatically to conserve the battery.
Alarm repeat: Unless the cause of an alarm is corrected within 2 minutes after the operator has pressed the “Silence” buzzer key, the buzzer will sound again to remind the operator of the problem.
Data memorize function: Retains the most recent flow rate, volume limit or total volume infused for the next use.
Sound setting: Audible indication of standby and infusion in progress can be turned on or off.
Power source switch alarm: An alarm sounds when AC power supply has stopped and the unit has switched over to battery operation..
 
Operating conditions
Ambient temperature: +5 to 40°C
Humidity: 20 to 90% RH (no condensation)
Pressure: 70 to 106kPa
 
Storage conditions
Ambient temperature: -10 to 45°C
Humidity: 10 to 90% (no condensation)
Pressure: 50 to 106kPa
 
Power Source
AC 100-240V±10%, 50/60 Hz
Internal Battery: Rechargeable DC 12V, 700mAh, Ni-Cd (type BP-53).
Battery life is approx. 2 hours when using a new, fully charged battery at a fluid delivery rate of 25 ml/h.
External DC power supply:P DC 12.7 to 15 V (external power cord optional)
 
Input current
AC power supply: 0.3 A~ External DC power supply: 0.7 A
 
Classification
Class I, Type CF, IPX1 (drip proof)
 
Fuse
T1A
 
Dimension
90(W) x 216(H) 209(D)mm
 
Weight
Approx. 2.6kg
 
Accessories
AC power cord, drop sensor, Operation guide
 
Options
Custom stand for TOP infusion pumps/Multi-pump stand/Cable for multi-pump connection/ Pole clamp/ External power cord/Cable for nurse call connection/ External communication cable.
 
TOP IV Sets H-series

TOP IV sets H-series have been designed specifically for use with TOP infusion pumps. The tubing inner diameter has been held to a closer tolerance than in general-use TOP IV sets and the tubing itself is made of a highly resilient material, enabling very stable infusion with TOP infusion pumps.